Step1:
Powering Up System
Step2: Draping Tray
Arm
Step3: Set Up Fluidics
> Step4:
Connect Handpiece and Prime Fluidics Set
Step5: Switching to
I/A Cannula
Step4:
Connect Handpiece and Prime Fluidics Set |
 |
4.1. Fully
insert Handpiece bipolar plugs into Bipolar outputs of High
Frequency Generator. |
 |
4.2. Lightly
connect irrigation line to irrigation pigtail of Handpiece.
(both have color strip) |
 |
4.3. With
cap in place, place Handpiece tip up. |
 |
4.4. Turn
Flow Control to “Prime” (8 o’clock position). |
 |
4.5. Partially
open Roller Clamp allowing BSS to fill irrigation line. Once
BSS emerges from tip of Handpiece, open Roller Clamp completely. |
 |
4.6. Once
cap is filled to the height of the tip or BSS emerges from
aspiration pig tail of Handpiece, reposition Handpiece tip
down (with cap in place). |
 |
4.7. Lightly
connect aspiration line. Continue to fill until BSS begins
to appear in collection bag and all lines are free of air
bubbles. |
 |
4.8. Turn
Flow Control to “Standby” until Foot Pedal is positioned
for surgery. |
 |
4.9. Turn
Flow Control to physician’s preferred surgical flow setting,
typically “Flow 3” (12 o’clock position). |
 |
4.10. Cover
and secure small drape over I/A Console and High Frequency
Generator and check that the irrigation tubing is not kinked. |
| |
Step5:
Switching to I/A Cannula > |